This Technical Specification describes a standard method for determining the particle number concentration in ambient air in a range up to about 107 cm-3 for averaging times equal to or larger than 1 min. The standard method is based on a Condensation Particle Counter (CPC) operated in the counting mode and an appropriate dilution system for concentrations exceeding the counting mode range. It also defines the performance characteristics and the minimum requirements of the instruments to be used. The lower and upper sizes considered within this document are 7 nm and a few micrometres, respectively. This document describes sampling, operation, data processing and QA/QC procedures including calibration parameters.
PD CEN/TS 16976:2016 Referenced Document
ISO 27891:2015 Aerosol particle number concentration - Calibration of condensation particle counters
PD CEN/TS 16976:2016 history
2016PD CEN/TS 16976:2016 Ambient air - Determination of the particle number concentration of atmospheric aerosol