CFR 21-803.40-2014 Food and Drugs. Part803:Medical device reporting (Eff. 8-14-15). Section803.40:If I am an importer, what reporting requirements apply to me?
(a) Reports of deaths or serious injuries. You must submit a report to us, and a copy of this report to the manufacturer, as soon as practicable, but no later than 30 calendar days after the day that you receive or otherwise become aware of information fr
CFR 21-803.40-2014 history
2014CFR 21-803.40-2014 Food and Drugs. Part803:Medical device reporting (Eff. 8-14-15). Section803.40:If I am an importer, what reporting requirements apply to me?