Scope

5.1 This test method is intended to be used by wire producers and thermocouple manufacturers for certification of refractory metal thermocouples. It is intended to provide a consistent method for calibration of refractory metal thermocouples referenced to a calibrated radiation thermometer. Uncertainty in calibration and operation of the radiation thermometer, and proper construction and use of the test furnace are of primary importance.

5.2 Calibration establishes the temperature-emf relationship for a particular thermocouple under a specific temperature and chemical environment. However, during high temperature calibration or application at elevated temperatures in vacuum, oxidizing, reducing or contaminating environments, and depending on temperature distribution, local irreversible changes may occur in the Seebeck Coefficient of one or both thermoelements. If the introduced inhomogeneities are significant, the emf from the thermocouple will depend on the distribution of temperature between the measuring and reference junctions.

5.3 At high temperatures, the accuracy of refractory metal thermocouples may be limited by electrical shunting errors through the ceramic insulators of the thermocouple assembly. This effect may be reduced by careful choice of the insulator material, but above approximately 21008201;°C, the electrical shunting errors may be significant even for the best insulators available.

1.1 This test method covers the calibration of refractory metal thermocouples using a radiation thermometer as the standard instrument. This test method is intended for use with types of thermocouples that cannot be exposed to an oxidizing atmosphere. These procedures are appropriate for thermocouple calibrations at temperatures above 8008201;°C (14728201;°F).

1.2 The calibration method is applicable to the following thermocouple assemblies:

1.2.1 Type 1—Bare-wire thermocouple assemblies in which vacuum or an inert or reducing gas is the only electrical insulating medium between the thermoelements.

1.2.2 Type 2—Assemblies in which loose fitting ceramic insulating pieces, such as single-bore or double-bore tubes, are placed over the thermoelements.

1.2.3 Type 2A—Assemblies in which loose fitting ceramic insulating pieces, such as single-bore or double-bore tubes, are placed over the thermoelements, permanently enclosed and sealed in a loose fitting metal or ceramic tube.

1.2.4 Type 3—Swaged assemblies in which a refractory insulating powder is compressed around the thermoelements and encased in a thin-walled tube or sheath made of a high melting point metal or alloy.

1.2.5 Type 4—Thermocouple assemblies in which one thermoelement is in the shape of a closed-end protection tube and the other thermoelement is a solid wire or rod that is coaxially supported inside the closed-end tube. The space between the two thermoelements can be filled with an inert or reducing gas, or with ceramic insulating materials, or kept under vacuum.

1.3 This standard does not purpor......

ASTM E452-02(2013) Referenced Document

  • ASTM E1256 Standard Test Methods for Radiation Thermometers (Single Waveband Type)
  • ASTM E1751 Standard Guide for Temperature Electromotive Force (emf) Tables for Non-Letter Designated Thermocouple Combinations
  • ASTM E344 Terminology Relating to Thermometry and Hydrometry
  • ASTM E563 Standard Practice for Preparation and Use of an Ice-Point Bath as a Reference Temperature
  • ASTM E988 Standard Temperature-Electromotive Force (EMF) Tables for Tungsten-Rhenium Thermocouples

ASTM E452-02(2013) history

  • 2023 ASTM E452-02(2023) Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer
  • 2018 ASTM E452-02(2018) Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer
  • 2002 ASTM E452-02(2013) Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer
  • 2002 ASTM E452-02(2007) Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer
  • 2002 ASTM E452-02 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer
  • 1997 ASTM E452-97 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer
Standard Test Method for  Calibration of Refractory Metal Thermocouples Using a Radiation  Thermometer



Copyright ©2024 All Rights Reserved