ASTM F1800-12
Standard Practice for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements

Standard No.
ASTM F1800-12
Release Date
2012
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM F1800-19
Latest
ASTM F1800-19e1
Scope

4.1 This practice can be used to describe the effects of materials, manufacturing, and design variables on the fatigue performance of metallic tibial trays subject to cyclic loading for relatively large numbers of cycles.

4.2 The loading of tibial tray designs in vivo will, in general, differ from the loading defined in this practice. The results obtained here cannot be used to directly predict in vivo performance. However, this practice is designed to allow for comparisons between the fatigue performance of different metallic tibial tray designs, when tested under similar conditions.

4.3 In order for fatigue data on tibial trays to be comparable, reproducible, and capable of being correlated among laboratories, it is essential that uniform procedures be established.

1.1 This practice covers a procedure for the fatigue testing of metallic tibial trays used in knee joint replacements. This practice covers the procedures for the performance of fatigue tests on metallic tibial components using a cyclic, constant-amplitude force. It applies to tibial trays which cover both the medial and lateral plateaus of the tibia. This practice may require modifications to accommodate other tibial tray designs.

1.2 This practice is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic tibial trays with one unsupported condyle.

1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ASTM F1800-12 Referenced Document

  • ASTM E1150 Definitions of Terms Relating to Fatigue
  • ASTM E4 Standard Practices for Force Verification of Testing Machines
  • ASTM E467 Standard Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System
  • ASTM E468 Standard Practice for Presentation of Constant Amplitude Fatigue Test Results for Metallic Materials

ASTM F1800-12 history

  • 2019 ASTM F1800-19e1 Standard test method for cyclic fatigue testing of medical tibial disc components of total knee replacements
  • 2019 ASTM F1800-19 Standard Practice for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
  • 2012 ASTM F1800-12 Standard Practice for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
  • 2007 ASTM F1800-07 Standard Test Method for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
  • 2004 ASTM F1800-04 Standard Test Method for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
  • 1997 ASTM F1800-97(2003) Standard Test Method for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
  • 1997 ASTM F1800-97 Standard Test Method for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements
Standard Practice for  Cyclic Fatigue Testing of Metal Tibial Tray Components of Total  Knee Joint Replacements



Copyright ©2024 All Rights Reserved