Turbidity is undesirable in drinking water, plant effluent waters, water for food and beverage processing, and for a large number of other water-dependent manufacturing processes. Removal is often accomplished by coagulation, settling, and filtration. Measurement of turbidity provides a rapid means of process control for when, how, and to what extent the water must be treated to meet specifications.
This test method is suitable to turbidity such as that found in drinking water, process water, and high purity industrial water.
When reporting the measured result, appropriate units should also be reported. The units are reflective of the technology used to generate the result, and if necessary, provide more adequate comparison to historical data sets.
Table 1 describes technologies and reporting results (see also Refs (1),(2),(3)). Those technologies listed are appropriate for the range of measurement prescribed in this method. Others may come available in the future. Fig. X5.1 provides a flow chart to aid in selection of the appropriate technology for low-level static turbidity applications.
If a design that falls outside of the criteria listed in Table 1 is used, the turbidity should be reported in turbidity units (TU) with a subscripted wavelength value to characterize the light source that was used.
TABLE 1 Applicable Technologies Available for Performing Static Turbidity Measurements Below 5 NTU
Design and Reporting Unit | Prominent Application | Key Design Features | Typical Instrument Range | Suggested Application |
Nephelometric non-ratio (NTU) | White light turbidimeters. Comply with USEPA Method 180.1 (1) for low level turbidity monitoring. | Detector centered at 90° relative to the incident light beam. Uses a white light spectral source. | 0.020 to 40 | Regulatory reporting of clean water |
Ratio White Light turbidimeters (NTRU) | Complies with ISWTR regulations and Standard Method 2130B. (2) Can be used for both low and high level measurement. | Used a white light spectral source. Primary detector centered at 90°. Other detectors located at other angles. An instrument algorithm uses a combination of detector readings to generate the turbidity reading. | 0.020 to10 000 | Regulatory Reporting of clean water |
Nephelometric, near-IR turbidimeters, non-ratiometric (FNU) | Complies with ISO 7027. The wavelength is less susceptible to color interferences. Applicable for samples with color and good for low level monitoring. | Detector centered at 90° relative to the incident light beam. Uses a near-IR (780-900 nm) monochromatic light source. | 0.012 to 1000 | 0 - 40 ISO 7027 Regulatory reporting |
Nephelometric near-IR turbidimeters......... ASTM D6855-12 Referenced Document
ASTM D6855-12 history
Copyright ©2024 All Rights Reserved |