ASTM E2603-08
Standard Practice for Calibration of Fixed-Cell Differential Scanning Calorimeters

Standard No.
ASTM E2603-08
Release Date
2008
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM E2603-15
Latest
ASTM E2603-15(2023)
Scope

Fixed-cell differential scanning calorimeters are used to determine the transition temperatures and energetics of materials in solution. For this information to be accepted with confidence in an absolute sense, temperature and heat calibration of the apparatus or comparison of the resulting data to that of known standard materials is required.

This practice is useful in calibrating the temperature and heat flow axes of fixed-cell differential scanning calorimeters.

1.1 1.1 This practice covers the calibration of fixed-cell differential scanning calorimeters over the temperature range from 10 to +120°C.

1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 7.