ASTM UOP678-04
Dissolved Molecular Oxygen in Liquid Hydrocarbons by Electrochemical Detection

Standard No.
ASTM UOP678-04
Release Date
2004
Published By
American Society for Testing and Materials (ASTM)
Latest
ASTM UOP678-04
Scope

This method is for estimating the dissolved oxygen in liquid hydrocarbons by electrochemical detection. The preferred application is online analysis, although the analysis may also be performed in the laboratory. The lower detection limit is 0.1 mass-ppm while the upper limit is dependent on the solubility of oxygen in the specific sample. Carbon dioxide, ammonia, hydrogen sulfide, and mercaptans at concentrations above 200, 300, 15 and 15-ppm (see Note 1) respectively, will interfere with the oxygen determination. Materials that contaminate the membrane or volatile components that are either oxidized, reduced, or react with the sensor electrodes will also interfere. The sampling system associated with a process unit is to be specified depending on the specific process unit.

ASTM UOP678-04 history

  • 2004 ASTM UOP678-04 Dissolved Molecular Oxygen in Liquid Hydrocarbons by Electrochemical Detection
  • 1988 ASTM UOP678-88 Dissolved Molecular Oxygen in Liquid Hydrocarbons by Electrochemical Detection



Copyright ©2024 All Rights Reserved