ASTM F78-97(2002)
Standard Test Method for Calibration of Helium Leak Detectors by Use of Secondary Standards

Standard No.
ASTM F78-97(2002)
Release Date
1997
Published By
American Society for Testing and Materials (ASTM)
Status
 2008-07
Latest
ASTM F78-97(2002)
Scope

1.1 This test method covers a procedure for calibrating a mass spectrometer-type helium leak detector with a series of commercially available calibrated leaks without need for recourse to a primary standard.

1.2 Leak detector parameters determined by this test method include:

1.2.1 Minimum detectable signal, drift noise (8.5, with recorder; 8.6, without recorder),

1.2.2 Response time,

1.2.3 Minimum detectable leak rate, and

1.2.4 Sensitivity.

1.3 This standard does not purport to address the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ASTM F78-97(2002) Referenced Document

  • ASTM E1 Standard Specification for ASTM Thermometers*1998-04-19 Update
  • ASTM E425 

ASTM F78-97(2002) history

  • 1997 ASTM F78-97(2002) Standard Test Method for Calibration of Helium Leak Detectors by Use of Secondary Standards
  • 1991 ASTM F78-97 Standard Test Method for Calibration of Helium Leak Detectors by Use of Secondary Standards
Standard Test Method for Calibration of Helium Leak Detectors by Use of Secondary Standards