1.1 This specification covers the requirements for metal-sheathed industrial platinum resistance thermometers (PRT''s) suitable for direct immersion temperature measurement. It applies to PRT''s with an average temperature coefficient of resistance between 0 and 100176C of 0.385%/176C and nominal resistance at 0176C of 100 [omega] or other specified value. This specification covers PRT''s suitable for all or part of the temperature range -200 to 650176C. The resistance-temperature relationship and tolerances are specified as well as physical, performance, and testing requirements.
1.2 The values of temperature in this specification are based on the International Temperature Scale of 1990 (ITS-90) .