This part of ISO 8655 specifies — metrological requirements, — maximum permissible errors, — requirements for marking and — information to be provided for users, for air-displacement (type A) and positive-displacement (type D) single-channel and multi-channel piston pipettes, complete with their selected tip(s) and any other essential, consumable parts, designed to deliver their specified nominal volume (Ex). NOTE General requirements and definitions of terms of piston-operated volumetric apparatus are given in ISO 8655-1. Conformity testing (type evaluation) of piston-operated volumetric apparatus is given in ISO 8655-6. Alternative test methods such as photometric and titrimelric methods will be the subject of a future Part 7 to ISO 8655. For all other tests (e.g. quality assurance by the supplier, analytical and measuring equipment quality assurance by the user) see ISO 8655-6 or alternative test methods. For safety requirements of electrically powered piston pipettes, see IEC 61010-1.
BS EN ISO 8655-2:2002 Referenced Document
ISO 3696 Water for analytical laboratory use; Specification and test methods
ISO 8655-1:2002 Piston-operated volumetric apparatus - Part 1: Terminology, general requirements and user recommendations
ISO 8655-6:2002 Piston-operated volumetric apparatus - Part 6: Gravimetric methods for the determination of measurement error
BS EN ISO 8655-2:2002 history
2022BS EN ISO 8655-2:2022 Piston-operated volumetric apparatus. Pipettes (ISO 8655-2:2022)
2002BS EN ISO 8655-2:2002 Piston-operated volumetric apparatus - Part 2:Piston pipettes