BS EN 61582:2006 Radiation protection instrumentation - In vivo counters - Classification, general requirements and test procedures for portable, transportable and installed equipment
This International Standard specifies the classification, general design requirements,
performance characteristics and test procedures for in vivo counting systems for detecting
trace amounts of radionuclides in the bodies of persons working in nuclear power plants,
laboratories and facilities handling radionuclides, and inhabitants living on territory which may
be contaminated by either naturally occurring or artificial radionuclides. The purpose is to
determine the dose equivalent to organs and the effective dose of internal radiation for the
whole body.
This standard is applicable both to equipment with spectroscopic capabilities and instruments
for rapid screening for gross internal contamination only.
This standard is applicable to instruments for the monitoring of certain critical organs (for
example, lungs, thyroid gland, etc.) as well as instruments for monitoring the whole body.
The standard applies to equipment for the measurement of the activity of gamma-emitting
radionuclides in humans in order to determine the committed dose equivalent due to internal
contamination in accordance with the recommendations of the ICRP 60 and ICRP 61.
The requirements of the standard are applicable to the installed apparatus, to vehiclemounted
equipment and to portable instruments. However, Annex B defines the additional
mechanical and environmental performance requirements and the additional testing required
for transportable and portable assemblies. The general and radiological requirements of all
types of in vivo counters are included in this standard.
Depending on the type of instrument and the organ to be checked, measurement geometry
may require the subject of the monitoring procedure to stand, sit, or lie.
The detection assembly includes one or more radiation detector. Normally, these are
shielded-scintillation or semi-conductor detectors. Where identification of the location of
contamination is required, the detectors may be collimated.
The measurement assembly includes functional units for the processing of signals from the
detection assembly as well as units for the display of the measured activity.
For the measurement of body mass a built-in weighing machine may be used.
The standard specifies general types, specific measuring characteristics, main test
procedures, electrical and mechanical characteristics Text deleted, as well as the
requirements related to background radiation of the environment.
This equipment is not intended for the determination of external contamination of the human
body or the clothing of personnel.
This standard is not applicable to equipment such as radiation detectors intended for
introduction into the human body.
BS EN 61582:2006 Referenced Document
IEC 60068-2-1 Environmental testing - Part 2-1: Tests - Test A: Cold*, 2007-03-01 Update
IEC 60068-2-2 Environmental testing - Part 2-2: Tests - Test B: Dry heat*, 2007-07-01 Update
IEC 60068-2-6 Environmental testing - Part 2-6: Tests - Test Fc: Vibration (sinusoidal)*, 2007-12-01 Update