BS EN 16479:2023 Water quality. Performance requirements and conformity test procedures for water monitoring equipment. Automatic sampling devices (samplers) for water and waste water
1 Scope
This document specifies general requirements, performance requirements and conformity
test procedures for automatic sampling devices (samplers) for water and waste water
that:
— sample water and waste water from non-pressurized (i.e. open to atmosphere) channels
or vessels;
— sample over extended periods to collect discrete or composite samples based on time, event or flow proportional sampling.
It does not include sampling systems built into online and in-line analysers.
The general requirements include functional facilities that samplers need to meet
users’ applications and information that needs to be included in associated documents.
The test procedures specify uniform methods to be used when determining key performance
characteristics of samplers at one or more set sample volume. It is for the sampler
manufacturer and/or user to decide on the required set sample volume(s). All of the
test procedures are to be carried out under laboratory conditions. It is recognized
that for some samplers, certain test procedures are not applicable.
Statistical procedures are specified for evaluation of the test data. Some example
calculations are provided.
Specific sample integrity requirements are specified for samplers to be used for the
collection of samples of final effluent or influent for the purpose of monitoring
the performance of urban waste w...
BS EN 16479:2023 history
2023BS EN 16479:2023 Water quality. Performance requirements and conformity test procedures for water monitoring equipment. Automatic sampling devices (samplers) for water and waste water
2014BS EN 16479:2014 Water quality. Performance requirements and conformity test procedures for water monitoring equipment. Automated sampling devices (samplers) for water and waste water