In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials." >

ASTM F2888-13
Standard Test Method for Platelet Leukocyte Count—An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials

Standard No.
ASTM F2888-13
Release Date
2013
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM F2888-19
Latest
ASTM F2888-19
Scope
1.1 This test method assists in the evaluation of cardiovascular device materials for their ability to induce thrombus formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and materials contacting human blood, as per ANSI/AAMI/ ISO 10993–4. See also Test Method F2382. 1.2 All safety policies and practices shall be observed during the performance of this test method. All human blood and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled as the contents, and disposed of by appropriate means. 1.3 The human blood should be handled at Biosafety Level 2 as recommended in the Centers for Disease Control/National Institutes of Health Manual Biosafety in Microbiological Laboratories. The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses. The blood should be treated like any patient blood in using universal precautions. 1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 7 on Hazards.

ASTM F2888-13 Referenced Document

  • ASTM F2382 Standard Test Method for Assessment of Circulating Blood-Contacting Medical Device Materials on Partial Thromboplastin Time (PTT)*2018-10-01 Update

ASTM F2888-13 history

  • 2019 ASTM F2888-19 Standard Practice for Platelet Leukocyte Count—An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials
  • 2013 ASTM F2888-13 Standard Test Method for Platelet Leukocyte Count—An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials
Standard Test Method for Platelet Leukocyte Count—An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials



Copyright ©2024 All Rights Reserved