This standard specifies the inductively coupled plasma mass spectrometry method for the determination of metallic elements in solid waste and solid waste leachate. This standard applies to silver (AG), arsenic (AS), barium (BA), beryllium (BE), cadmium (CD), cobalt (CO), chromium (CR), copper (CU), Determination of 17 metal elements including manganese (MN), molybdenum (MO), nickel (NI), lead (PB), antimony (SB), selenium (SE), thallium (TL), vanadium (V), and zinc (ZN) . If verified, this standard can also be applied to the determination of other metallic elements. When the sampling volume of solid waste leachate is 25 ML, the detection limits of 17 metal elements are 0.7 MG/L~6.4 MG/L, and the lower limit of determination is 2.8 MG/L~25.6 MG/L. Please see Appendix A for the method detection limits of each element. When the solid waste sample amount is 0.1 G, the method detection limit of 17 metal elements is 0.4 MG/KG ~ 3.2 MG/KG, and the lower determination limit is 1.6 MG/KG ~ 12.8 MG/KG.
HJ 766-2015 Referenced Document
HJ 557 Solid waste.Extraction procedure for leaching toxicity.Horizontal vibration method
HJ/T 20 Technical specifications on sampling and sample preparation from industry solid waste
HJ/T 298 Technical specifications on identification for hazardous waste