T/CHES 27-2019
Water Quality Determination of Ammonia Nitrogen Automatic Analysis Nessler's Reagent Spectrophotometry (English Version)

Standard No.
T/CHES 27-2019
Language
Chinese, Available in English version
Release Date
2019
Published By
Group Standards of the People's Republic of China
Latest
T/CHES 27-2019
Scope
This standard specifies the automated Nessler reagent spectrophotometric method for the determination of ammonia nitrogen in water. This standard is applicable to the determination of ammonia nitrogen in surface water, groundwater, drinking water, domestic sewage and industrial wastewater. When direct injection, the detection limit of this method is 0.02mg/L, and the detection range is 0.08mg/L~2.00mg/L; when distillation injection, the detection limit of this method is 0.05mg/L, and the detection range is 0.20mg /L~8.00mg/L. When the sample concentration exceeds the measurement range, it should be measured after dilution.

T/CHES 27-2019 history

  • 2019 T/CHES 27-2019 Water Quality Determination of Ammonia Nitrogen Automatic Analysis Nessler's Reagent Spectrophotometry