This standard specifies the graphite furnace atomic absorption spectrophotometric method for the determination of total chromium in solid waste and solid waste leach solutions. This standard is applicable to the determination of total waste in recycled waste and solid waste leachate. When measuring solid waste, the sample volume is 0.2 g, the fixed volume is S0 ml, and the injection volume is 20, the detection limit of this method is 0.2 mgkg, and the lower determination limit is 0.8 mg/kg. For the determination of solid waste leachate, when the constant volume is 50 ml and the input volume is 20 times, the detection limit of this method is 0.7 ngL, and the lower limit of determination is 2.8 ng/L.
HJ 750-2015 Referenced Document
GB 5086.1 Test method standard for leaching toxicity of solid wastes--Roll over leaching procedure
HJ/T 20 Technical specifications on sampling and sample preparation from industry solid waste
HJ/T 298 Technical specifications on identification for hazardous waste