This standard specifies the inductively coupled plasma emission spectrometry method for the determination of 32 elements in water. This standard applies to silver, aluminum, arsenic, boron, barium, beryllium, bismuth, calcium, cadmium, cobalt, chromium, copper, iron, potassium, lithium, magnesium, manganese, molybdenum, Determination of soluble elements and total elements of 32 elements including sodium, nickel, phosphorus, lead, sulfur, antimony, selenium, silicon, tin, strontium, titanium, vanadium, zinc and zirconium. The method detection limit of each element in this standard is 0.009 mg/L~0.1 mg/L, and the lower determination limit is 0.036 mg/L~0.39 mg/L. Please see Appendix A for the method detection limits of each element.
HJ 776-2015 Referenced Document
GB/T 6682 Water for analytical laboratory use.Specification and test methods
HJ 678 Water quality.Digestion of total metals.Microwave assisted acid digestion method
HJ/T 164 Technical specifications for environmental monitoring of groundwater
HJ/T 91 Technical Specifications Requirements for Monitoring of Surface Water and Waste Water
HJ 776-2015 history
2016HJ 776-2015 Water quality – Determination of 32 elements-Inductively coupled plasma optical emission spectrometry